EcoGen 320
Cost:
On request
Description
Specifications
The EcoGen line of stationary X-ray devices of the cable type is designed to solve a wide range of tasks in radiography and fluoroscopy. Provide high-quality and reliable generation of X-ray radiation.
The standard EcoGen 320 kit includes the following components:
- cathode high voltage generator
- anode high voltage generator
- ceramic metal X-ray tube
- high–voltage cable – 2 pcs.
- X-ray machine control panel
- closed-loop oil cooling system
- cable set
It is possible to retrofit the kit with additional equipment on request.
The main features of stationary X-ray devices of the EcoGen cable type:
- High-quality long-term radiation, with low pulsations of high voltage and current, as well as high stability of parameters.
- The versatility of the equipment allows it to be used not only in industry, but also in scientific laboratories.
- The possibility of round-the-clock operation due to the safety margin of the generator in terms of power. In addition, this is facilitated by constant monitoring of the state of the device.
- Ease of installation, which allows not only to quickly put into operation, but also to integrate into existing industrial systems.
- Modular design of the X-ray machine.
Allows the replacement of individual components, simplifies the repair and retrofitting of the X-ray machine (for example, an additional X-ray tube, a longer high-voltage cable).
- A library of X-ray tubes and an automated procedure for their training.
- Maintainability. The high-voltage generator is made in the form of a single module consisting of a power module and a high-voltage unit. The high-voltage part is placed in oil, disassembled and assembled without degassing and vacuuming, which facilitates maintenance.
- An autonomous control panel with a graphical built-in monitor. User-friendly interface in English.
- All components of the X-ray machine pass the output test control at the factory.
EcoGen 320 high voltage generator
| Output voltage | 10 ÷ 320 kV |
| Output current | 0 ÷ 45 mA |
| Maximum generator power | 4500 W |
| Accuracy of setting the output voltage | < ±1% |
| Reproducibility of output voltage | ±0,01% at constant temperature |
| Accuracy of setting the output current | ±0,1% |
| Reproducibility of output current | ±0,01% at constant temperature |
| Power supply | 220V/380V |
| Frequency | 50Hz |
Bipolar tubes of the EcoGen 320 X-ray machine
| 320-04.10 | 320-30.55 | 320-19.36 | 320-80 | |
| Maximum voltage | 320kV | |||
| Focal spot size IEC336 | 0,15mm | 1,5mm | 0,8mm | |
| 0,4mm | 3,0mm | 1,8mm | 5,0mm | |
| Focal spot size EN12543 | 0,4mm | 3,0mm | 1,9mm | |
| 1,0mm | 5,5mm | 3,6mm | 8,0mm | |
| Maximum power of the tube at the anode | 800W | 1500W | 640W | |
| 1800W | 4200W | 1600W | 4000W | |
| Tube current at max. tube voltage | 10mA | 14mA | 95mA | |
| 22mA | 46mA | 20mA | 60mA | |
| Radiation output angle | 40°х30° | 40°х40° | 40°х40° | 40°х40° |
| Type of high voltage connector | R28 | |||
| Weight | 40kg | 40kg | 40kg | 40kg |